Specifications
| Connection Inlet Flange | DN 200 CF (10") |
| Connection Outlet Flange | DN 25 ISO-KF (NW25) |
| Cooling Method | Air Cooling (convection) Air Cooling (forced - fan) Water Cooling |
| Power Supply Voltage | DC 48V |
| Power Total Rated | 600.00 |
| Pumping Speed Air Nitrogen (@50Hz or max @default setting) | 0.850 |
| Pressure Ultimate | 3.000 × 10⁻⁸ |
TURBOVAC 950 i - DN 200 CF
Part Number: 840111V1000
Variant
Explore the range of variants given below to compare the product
| Short Description | Part Number | Variant TMP |
|---|---|---|
| 840091V1000 | TURBOVAC i - frontend 1 (basic) | |
| 840091V2000 | TURBOVAC i - frontend 2 (Anybus/Fieldbus extention) | |
| 840091V3000 | TURBOVAC iX - Xtention box controller | |
| 840091V3300 | TURBOVAC iX - Xtention box controller | |
| 840111V1000 | TURBOVAC i - frontend 1 (basic) | |
| 840111V2000 | TURBOVAC i - frontend 2 (Anybus/Fieldbus extention) | |
| 840111V3000 | TURBOVAC iX - Xtention box controller | |
| 840111V3300 | TURBOVAC iX - Xtention box controller |
General Product Information
| Accessories / Scope of Supply | TURBOVAC i pumps come stand-alone - the suitable power supply unit, comms/power cables, and cooling options need to be ordered separately. IND industrial variants come with an equipped water cooling unit and are water cooling only. |
| Connection Cooling |
6x1 mm flexible hose quick connection
1/8" BSPP (G) Female |
| Connection Outlet Flange | DN 25 ISO-KF (NW25) |
| Connection Inlet Flange | DN 200 CF (10") |
| Cooling Method |
Air Cooling (convection)
Air Cooling (forced - fan) Water Cooling |
| Dimension Height | 0.307 |
| Dimension Length | 0.261 |
| Dimension Width | 0.262 |
| IP Rating | IP40 |
| Noise / Sound Level |
50.00
db(A)
50.00 db(A) |
| Power Supply Frequency | DC Voltage |
| Power Supply Voltage | DC 48V |
| Weight | 21.70 |
| Dimension Comments | The dimensions length, width and height try to describe the outer shape of this particular product in its typical orientation when standing on the ground. We always recommend checking the detailed dimensions given in 2D drawings in the instruction manual or the 3D shapes given in 3D models. |
| Connection Inlet / Outlet Flange | DN 200 CF (10") | DN 25 ISO-KF (NW25) |
| Certification of Product |
CE
UL (61010-1 3rd Edition) CSA (C22.2 No.61010-1-12) |
| Connection Purge Gas, Gas Ballast, Vent |
1/8" BSPP (or G 1/8") Female
1/8" BSPP (or G 1/8") Female |
Environmental Conditions
| Temperature Ambient Operation Max | 40.00 |
| Temperature Ambient Storage Max | 70.00 |
| Temperature Ambient Operation Min | 5.00 |
| Temperature Ambient Storage Min | -15.00 |
| Magnetic Field Resistance Max (covering worst direction - come and consult us for more details) | 0.010 T |
Operation Conditions - Limitations
| Altitude Max | 4000.00 |
| Flow Cooling Water Min | 1.667 × 10⁻⁵ |
| Temperature Cooling Water Max | 35.00 |
| Temperature Cooling Water Min | 15.00 |
| Flow Cooling Water Max | 3.000 × 10⁻⁵ |
| Pressure Permissible Supply Cooling Water - Min and Max |
4.000 × 10⁵
4.000 × 10⁵ |
Classification
| Classification Dry or Wet Pump | Dry |
| Classification Product Pressure Range |
High Vacuum: 1·10⁻³ - 1·10⁻⁷ mbar
Ultra High Vacuum: 1·10⁻⁷ - 1·10⁻¹¹ mbar |
| Classification High / Medium / Rough Vacuum pump | High Vacuum Pump |
Performance Data
| Compression Ratio Air Nitrogen | 1.000 × 10¹¹ |
| Compression Ratio Helium | 1.000 × 10⁸ |
| Compression Ratio Hydrogen | 4.000 × 10⁶ |
| Power Total Rated | 600.00 |
| Pressure Ultimate | 3.000 × 10⁻⁸ |
| Pumping Speed Air Nitrogen (@50Hz or max @default setting) | 0.850 |
| Pumping Speed Argon | 0.790 |
| Pumping Speed Helium | 0.860 |
| Pumping Speed Hydrogen | 0.720 |
| Ramp Up Time | 150.00 |
| Nominal rotational speed | 820.00 |
| Performance Comment | The compression of a TMP is a value describing how good the design can compress a gas type at special conditions. Compression data is measured only using the CF flange variants. The values are stated for all family members. For some pump classes we are stating different pumping speeds ISO > CF flange as small differences in geometry influence their performance leading to a difference >5%. |
| Gas Throughput Air Nitrogen Max | 1.40 |
| Gas Throughput Helium Max | 2.00 |
| Gas Throughput Max Hydrogen | 2.00 |
| Gas Throughput Argon Max | 0.350 |
| Pressure Backing Max | 1500.00 |
| Compression Ratio Argon | 1.000 × 10¹¹ |
Connectivity - Communication Interface
| Communication Interface Protocol | USS Universal Serial Interface Protocol |
| Communication Interfaces Customer |
RS485 (serial)
REMOTE X1 (Parallel I/O) USB |
| Connection Electrical Comms Interface Connector |
DSUB 9-PIN - Male
DSUB HD 15-PIN - Female (high density) USB type B - Female |
Variants
| Installation Orientation | Any orientation |
| Variant TMP | TURBOVAC i - frontend 1 (basic) |
- English - Instructions - TURBOVAC 850i, 950i, 1350i, 1450i Bearing Exchange Kit
- English - Instructions - TURBOVAC i Var 1 Serial Interfaces
- English - Instructions - TURBOVAC i(X) 850 i(X), 950 i(X), 1350 i, 1450 i
- English - Sales Material - TURBOVAC i/iX brochure
- English - Sales Material - TURBOVACi/iX brochure
- English, German - Sales Material - Elevator Pitch Turbovac (EN_DE)
- English, German - Sales Material - Elevator Pitch TurbovacIX (EN_DE)